Benzene, 0.15 - 10 ppm (8610030)
Benzene, 0.15 - 10 ppm Part # 8610030
Measuring Range:
0.15 to 10 ppm
Typical Measuring Time:
approx. 25 to 150 s
Accuracy:
± 25 % (under calibration conditions)
Ambient Conditions:
Temperature:
0 to 30 °C
Temperature correction:
20 to 30 °C none
0 to 19 °C + 4% from the measured value per °C
Humidity:
1 to 25 mg/L
Humidity correction:
none
Cross Sensitivity:
No influence at 1 ppm benzene by: 50 ppm toluene 50 ppm xylene 800 ppm n-octane
Hose and probe measurement
The following flushing time recommendations for hose and probe measurements are marked with a brand-new, dry and clean hose and probes
Hose Length (meter) Flashing time (min)
FKM Hose 5-20 3
FKM hose 45 5
PTFE-lined PVC hose 3-15 7
Bar probe 90 (83 16 532) 1
Telescopic probe ES 150 (83 16 533) 2
Probe Set GP600 (83 28 667) 3