Formaldehyde, 0.15 - 3 ppm (8610100)
Formaldehyde, 0.15 - 3 ppm, Part # 8610100
Measuring Range:
0.15 to 3 ppm
Typical Measuring Time:
300 to 600 s
Accuracy:
± 25 % (under calibration conditions)
Ambient Conditions:
Temperature:
0 to 40 °C
Temperature correction:
none
Humidity:
4 to 20 mg/L
Humidity correction:
< 8 mg/L +40 % from measurement value per mg/L
8 to 12 mg/L none
> 12 mg/L -10 % from measurement value per mg/L
Cross Sensitivity:
No influence at 0.6 ppm formaldehyde by: 0.25 ppm acrolein 10 ppm vinyl acetate 50 ppm acetaldehyde 100 ppm styrene
Hose and probe measurement
The following flushing time recommendations for hose and probe measurements are marked with a brand-new, dry and clean hose and probes.
Hose Length (meter) Flashing time (min)
FKM Hose 5-20 3
FKM hose 45 5
PTFE-lined PVC hose 3-15 3
Bar probe 90 (83 16 532) 1
Telescopic probe ES 150 (83 16 533) 2
Probe Set GP600 (83 28 667) 3