Chromic Acid 0.1/a, 0.1 - 0.5 mg/m3, (6728681)
Chromic Acid 0.1/a, 0.1 - 0.5 mg/m^3, Part # 6728681, 9 Tests
Application Range
Standard Measuring Range: 0.1 to 0.5 mg/m3 Discoloration compared to color standard.
Number of Strokes n: 40
Time for Measurement: approx. 8 min
Standard Deviation: ± 50 %
Color Change: white violet
Ambient Operating Conditions
Temperature: 5 to 40 °C
Absolute Humidity: < 20 mg H2O / L
Reaction Principle
a) CrO3 + H2SO4 CrVI
b) CrVI + diphenylcarbazide CrIII + diphenylcarbazone
Cross Sensitivity
Metal chromates such as zinc or strontium chromate are indicated with about half the sensitivity. CrIII compounds do not affect the indication. Very high chromate concentrations lead to a rapid bleaching of the indication. Measurements should be repeated with less strokes.
Additional Information
After carrying out the required 40 pump strokes the reagent ampoule must be broken, the liquid transferred onto the indicating layer and carefully drawn through it using the pump