1,3-Butadiene, 0.5 - 25 ppm (8610300)
1,3-Butadiene, 0.5 - 25 ppm Part # 8610300
Measuring Range:
0.5 to 25 ppm
Typical Measuring Time:
10 to 280 s
Accuracy:
± 15 % (under calibration conditions)
Ambient Conditions:
Temperature:
0 to 40 °C
Temperature correction:
0 to 25 °C none
> 25 °C - 2.5 % from measurement value per °C
Humidity:
5 to 30 mg/L
Humidity correction:
< 10 mg/L 8 % from measurement value per mg/L
> 10 mg/L→ - 1 % from measurement value per mg/L
Cross Sensitivity:
No influence at 2 ppm 1,3 butadiene by: 5 ppm H2S 25 ppm white spirit 50 ppm toluene 50 ppm n-hexane 100 ppm benzene
Hose and probe measurement
The following flushing time recommendations for hose and probe measurements are marked with a brand-new, dry and clean hose and probes.
Hose Length (meter) Flashing time (min)
FKM Hose 5-20 3
FKM hose 45 5
PTFE-lined PVC hose 3-15 3
Bar probe 90 (83 16 532) 1
Telescopic probe ES 150 (83 16 533) 1
Probe Set GP600 (83 28 667) 3