The innovative Dräger X-act® 7000 analysis system consists of Dräger MicroTubes and an opto-electronic analysis device that lets you precisely measure gases in the low ppb range. It gives you precise results right on site, replacing slow, expensive lab analyses. It's extremely easy to use: insert the Dräger MicroTubes, start measuring, then read out the test result.
Benefits
Measures the lowest-quantity concentrations with lab-quality analysis Concentrations of hazardous substances in ambient workplace air must never exceed the defined limits. Monitoring these sometimes very low concentration levels is a responsible task. The Dräger X-act 7000 in combination with Dräger MicroTubes is focused on measuring carcinogenic and toxic substances in the low ppb range. This sensitive testing system is based on colorimetric chemical sensors capable of replacing conventional laboratory analysis and providing precise results right on site. This saves time and money.
More safety thanks to high selectivity The substance-specific reactive layers and multiple pre-layers in the Dräger MicroTubes allow for selective gas testing. The various layers serve as a filter. During testing, they filter out other substances that occur in the workplace so that only the target substance is reflected in the test result. This significantly reduces crosssensitivities and the number of false-positive test results.
Explosion-proof and ready to use The RFID tags applied to the Dräger MicroTubes contain all the calibration data that apply throughout their normal lifespan of one year. This eliminates the need for lengthy function tests and manual calibration. Because it works on the basis of mass-flow measurement, air pressure fluctuations have no influence on the device. Any possible temperature or humidity influences are already accounted for during the factory calibration. The analyzer is explosion-proof and certified under ATEX and CSA. The system is also dust- and splash-resistant in accordance with IP54. It also meets the requirements for electromagnetic compatibility under EN 61326-1.
Easy to use After an automatic self-test, the X-act 7000 analysis system is immediately ready for use. It is suitable for testing with all available Dräger MicroTubes. Simply insert the right Dräger MicroTube for your testing needs. The automatic drive carefully draws the MicroTube in and positions it. Control your test using the 3-button control and 2.4 inch colour display. When the test is complete, the green LED will glow and a message is displayed on the screen. You can store the test result, location and time in the internal data logger and read it later using the Dräger CC Vision software. The device is powered by five easily replaceable AA alkaline batteries. The battery life lasts for more than ten hours of testing and can be read on the display.
Also suitable for remote sampling Using the coupler (connector piece), the Dräger X-am® Pump can be attached to the X-act 7000. This allows you to test for carcinogenic and toxic substances in the ppb range, even in hard-to-reach places like ducts, pipelines or tanks up to a distance of 45 meters (147.6 feet). Because the X-am Pump is also certified explosion-proof for Zone 0, it is perfectly suited for these tasks.
Technical Data
Measuring range and resolution
Varies, depending on the Dräger MicroTubes in use
Typical testing time
From 5 seconds to 1,200 seconds, depending on the Dräger MicroTubes in use and the concentration of the test substance.
Ready to use Immediately
Calibration Not necessary
Operating temperature
0 to +50 °C (+32 to +122 °F)
Storage temperature
-20 to +55 °C (-4 to +131 °F) (Analyzer) +4 to +25 °C (+39 to +77 °F) (Dräger MicroTubes)
Air pressure
700 to 1,300 hPa
Humidity
0 to 95 % relative humidity, non-condensing
Recording of measured data
CMOS sensor optics and LED light
System diagnostics
Automatic, using micro-controllers for all system components
Display
2.4 inch high contrast, color display with a resolution of 240 x 320, viewing angle of 160°
Menu languages
30
Operating time
Approx. 600 minutes of testing time
Power supply
5 x 1.5 v AA alkaline batteries
Dimensions (L x W x H)
195 mm x 85 mm x 70 mm (7.7 in x 3.3 in x 2.8 in)
Approvals
ATEX/IECEx II 1G Ex IIC T4 Ga cCSAus Class I, Zone 0, AEx ia IIC T4 Ga
Protection class
IP54, dust- and splash-resistant
Dräger MicroTubes
Benzene 8610600
Test Range - 1 – 150 ppb
Test Time - 100 – 900 seconds
Benzene 8610030
Test Range - 0.15 – 10 ppm
Test Time - 25 – 150 seconds
1.3-Butadiene 8610460
Test Range - 25 – 500 ppb
Test Time - 200 – 600 seconds
1.3-Butadiene 8610300
Test Range - 0.5 – 25 ppm
Test Time - 10 – 280 seconds
Ethylene oxide 8610200
Test Range - 25 – 250 ppb
Test Time - 150 – 600 seconds
Ethylene oxide 8610580
Test Range - 0.25 – 10 ppm
Test Time - 100 – 500 seconds
Formaldehyde 8610540
Test Range - 5 – 150 ppb
Test Time - 360 – 960 seconds
Formaldehyde 8610100
Test Range - 0.15 – 3 ppm
Test Time - 300 – 600 seconds
Mercury 8610350
Test Range - 0.005 – 0.25 mg/m3
Test Time - 240 – 1,200 seconds