Carbon Monoxide 8/a, 8 - 150 PPM, (CH19701)
Carbon Monoxide 8/a, 8 - 150 PPM, Part # CH19701, 10 Tests
Application Range
Standard Measuring Range: 8 to 150 ppm
Number of Strokes n: 10
Time for Measurement: approx. 2 min
Standard Deviation: ± 10 to 15 %
Color Change: white pale brown
Ambient Operating Conditions
Temperature: 0 to 50 °C
Absolute Humidity: < 50 mg H2O / L
Reaction Principle
5 CO + I2O5 I + 5 CO2
Cross Sensitivity
Acetylene is also indicated, however, with less sensitivity. Petroleum hydrocarbons, benzene, halogenated hydrocarbons and hydrogen sulfide are retained in the pre-layer. In the case of higher concentrations of interfering hydrocarbons, use should be made of a carbon pre-tube (CH 24 101). Higher concentrations of easily cleavable halogenated hydrocarbons (e.g. trichloroethylene), are liable to form chromyl chloride in the pre-layer which changes the indicating layer to a yellowish-brown. CO determination is impossible in the case of high olefin concentrations