Hexane 10/a, 10 - 200 PPM / 300 - 2,500 PPM, (8103681)
Hexane 10/a, 10 - 200 PPM / 300 - 2,500 PPM, Part # 8103681, 10 Tests
Application Range
Standard Measuring Range: 10 to 200 ppm / 300 to 2500 ppm
Number of Strokes n: 5 / 1
Time for Measurement: approx. 75 s / approx. 15 s
Standard Deviation: ± 15 to 20 %
Color Change: white brown-green
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: 1 to 35 mg H2O / L
Reaction Principle
C5H12 + I2O5 I2
Cross Sensitivity
Many petrol hydrocarbons also are indicated, however, all with different sensitivitie. Differentiation is not possible. Aromatic hydrocarbons are only indicated with very low sensitivity. Carbon monoxide is indicated with slighlty lower sensitivity than n-hexane.
Extension of the Measuring Range
Using n = 11, divide the reading by 2; the measuring range will be 50 to 1,500 ppm.