Horiba U-52 Water Quality Meter Rental
Horiba U-52 Water Quality Meter
U-52 multi-parameter water quality checker enables to measure and indicate the monitoring result simultaneously up to 11 parameters with one unit. Tough & powerful design, considering user friendly field monitoring on site, such as rivers, ground water, drainage water, etc. bring together feature of function and performance, which makes simple operation for everybody.
Horiba U-52 Control Unit
- Auto-calibration feature provides hassle free calibration of pH, dissolved oxygen, conductivity, turbidity and depth.
- Shock resistant cover designed for rough treatment in the field and is easily cleaned.
- Cable can be easily connected and disconnected with quick-connect fitting.
Data Management for Horiba U-52
- Auto hold function freezes average data values on the screen to offer more time to verify or transcribe data.
- Diagnostic functions notify the user of errors.
- Integral USB connection for data transfer to a PC. USA cable is sold separately and includes software.
- Selectable measurement units allow the operator to report data without the need to convert data to desired units of measure.
Multiple Sensors Housed in Each Probe
- Multiple sensors allow for the measurement of 11 parameters simultaneously (pH, pH(mv), ORP, DO, COND, Salinity, TDS, Seawater Specific Gravity, Temperature, Turbidity, Water Depth).
- Improved stability of the dissolved oxygen sensor has been achieved with a new 3 electrode design for fast response and polarographic sensor for ease of maintenance.
- pH and ORP electrodes can be replaced individually to reduce replacement costs
- Measurement Principle: Glass Electrode Method
- Range: pH 0-14
- Resolution: 0.01 pH
- Repeatability: +/- 0.05 pH
- Accuracy: +/- 0.1 pH
Oxidation Reduction Potential (ORP)
- Measurement Principal: Platinum Electrode Method
- Range: -2000 mV to +2000 mV
- Resolution: 1 mV
- Repeatability: +/- 5 mV
- Accuracy: +/- 15 mV
Dissolved Oxygen (DO)
- Measurement Principal: Polarographic Method
- Range: 0 to 50.0 mg/L
- Resolution: 0.01 mg/L
- Repeatability: +/- 0.1 mg/L
- Accuracy: 0 to 20 mg/L: +/- 0.2 mg/L, 20 to 50 mg/L: +/- 0.5 mg/L
Conductivity (COND)
- Measurement Principal: 4 AC Electrode Method
- Range: 0 to 10 S/m (0 to 100 mS/cm)
- Resolution: 0.000 to 0.999 mS/cm: 0.001 mS/cm, 1.00 to 9.99 mS/cm: 0.01 mS/cm, 10.0 to 99.9 mS/cm: 0.1 mS/cm, 0.0 to 99.9 mS/m: 0.1 mS/m, 0.100 to 0.999 S/m: 0.001 S/m, 1.00 to 9.99 S/m: 0.01 S/m
- Repeatability: +/- .05% Full Scale (FS)
- Accuracy: +/- 1% Full Scale (FS) (Median of two-point calibration)
Salinity
- Measurement Principal: Conductivity Conversion
- Range: 0 to 70 PPT (permillage)
- Resolution: 0.1 PPT
- Repeatability: +/- 1 PPT
- Accuracy: +/- 3 PPT
Total Dissolved Solids (TDS)
- Measurement Principal: Conductivity Conversion
- Range: 0 to 100 g/L
- Resolution: 0.1% Full Scale
- Repeatability: +/- 2 g/L
- Accuracy: +/- 5 g/L
Seawater Specific Gravity
- Measurement Principal: Conductivity Conversion
- Range: 0 to 50
- Resolution: 0.1
- Repeatability: +/- 2
- Accuracy: +/- 5
Temperature
- Measurement Principal: Thermistor Method
- Range: -10 to 55° C
- Resolution: 0.01° C
- Repeatability: +/- 0.10° C (at Calibration Point)
- Accuracy: JIS Class B Platinum Thermometer Sensor (+/- 0.3 +0.005)
Turbidity (TURB)
- Measurement Principal: LED Light Source and 30 ° Scattering Method
- Range: 0 to 800 NTU
- Resolution: 0 to 99.9 NTU: 0.1 NTU, 100 to 800 NTU: 1 NTU
- Repeatability: +/- 5% (Reading) or +/- 0.5 NTU Whichever is Greater
- Accuracy: +/- 5% (Reading) or +/- 1 NTU Whichever is Greater
Sensor Probe
- Measurement Temperature: -10 to 55° C
- Maximum Sensor Diameter: Approx. 96 mm
- Probe Length: Approx. 340 mm
- Cable Length: Standard: 2 m, Optional: 10 or 30 m
- Mass: Approx: 1,800 g (Approx 3.97 lbs)
- Automatic Calibration: Yes, uses 4 pH Auto Cal Solution (1 point per sensor)
- Liquid Contact Part Material (Liquid and Material): PPS, SUS316L, SUS304, FKM, PEEK,Q, Titanium, FEP membrane, POM
- Water Resistance: JIS Protection Level 8
Control Unit
- Outer Dimensions: 115 (w) X 65 (D) X 283 (H) mm
- Mass: Approx. 800 g (Approx. 1.76 lbs)
- LCD: 320 X 240 Liquid Crystal Display w/ Backlight (Black & White)
- Data Memory: 10,000
- Communication: USB
- Battery: (4) C Batteries
- Water Resistance: JIS Protection Level 7 (When Sensor Cable is Fitted)
- Battery Life: Approx. 70 Hours (Without Backlight). *U-53 Approx. 500 Measurements
- Storage Temperature: -10 to 60° C
- Ambient Temperature: -5 to 45° C
Rental Unit Includes: Meter with 2m cable, display, flow cell, flow cell fitting kit, DO rebuild kit, pH reference solution, spare batteries, calibration cups, probe guard, 500ml of auto-cal solution, manual, download cable, software & software manual.