Phosphine 0.1/c, 0.5 - 3 PPM / 0.1 - 1.0 PPM, (8103711)
Phosphine 0.1/c, 0.5 - 3 PPM / 0.1 - 1.0 PPM, Part # 8103711, 10 Tests
Application Range
Standard Measuring Range: 0.5 to 3 ppm / 0.1 to 1.0 ppm
Number of Strokes n: 1 / 3
Time for Measurement: approx. 1 min / approx. 2.5 min
Standard Deviation: ± 10 to 15 %
Color Change: yellow red
Ambient Operating Conditions
Temperature: 2 to 40 °C
Absolute Humidity: < 40 mg H2O / L
Reaction Principle
HgCI2 + PH3 Hg-phosphide + HCI HCI + pH-Indicator red reaction product
Cross Sensitivity
A maximum of 6 ppm sulphur dioxide or 15 ppm hydrogen chloride have no influence on the reasing. Higher concentration result in plus errors. Ammonia (>100 ppm) result in minus errors. Hyrogen sulphide and Arsine are indicated with different sensitivity. 30 ppm hydrocyanic acid does not affect.